7
Troubleshooting
The following errors can cause faults.
Fault
The milling machine does not start
when the foot control is operated
Halogen light does not function
Vertical movement not smooth
If faults cannot be remedied using this fault finding chart, call your service technician or call the Dentaurum
customer services directly.
Tel.
+ 49 72 31 / 803 - 211
8
Spare parts
Mains cable
Fuse T 630 mA
Fuse F 4 A
Halogen light 12V – 5W
Chuck ø 2,35 mm
Chuck ø 3,00 mm
Milling oil
Instructions for use and EC Declaration of Conformity
Cause
Not connected to the mains
Control unit is not switched on
Foot control is not connected
Micromotor is not connected
Faulty fuse
Faulty milling head
Faulty halogen light bulb
Faulty fuse
Dirt in the columns
REF
907-027-10
907-032-00
907-999-00
907-647-00
094-222-20
Remedy
Insert the plug on the
control unit (illus. 3, 3.2)
Switch the control unit on at the
switched on ON/OFF switch (illus. 3, 3.1)
Insert the connecting cable of the
connected foot control into the socket
on the control unit (illus. 3, 3.8) and
screw on tightly
Insert the connecting cable of the
connected micromotor into the socket
on the spindle column and control unit
(illus. 3, 3.7) and screw on tightly
Change the fuse in the control unit
(illus. 7).
Refer to section 6
Inform Dentaurum customer services
Change halogen light bulb
Refer to section 6
Change the fuse in the control unit
(illus. 7).
Refer to section 6
Clean the columns with a brush
EN
33