Connecting The Sensors; Fig. 12 Connect The Sensors - WEBTEC HPM7000 Manual De Instrucciones

Ocultar thumbs Ver también para HPM7000:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Starting Up
5.3

Connecting the Sensors

Before the device can be used to take measurements, the sensors
required for the measurement must be connected.
Fig. 12
Connect the sensors
Connect the sensors in the following sequence:
1 Connect the sensors on the application side (e.g. hydraulic
connection).
2 Connect the sensors to a bus cable or a Y-junction (refer to figure).
3 Connect a bus cable to the corresponding port on the device which
is switched on.
IMPORTANT
The device does not record data from sensors which are connected
while a measurement is in progress.
Restart the measurements in order to record data from sensors just
connected.
Measuring data from sensors which are disconnected while a mea-
surement is in progress is recorded by the device to the moment of
disconnection.
34
Tabla de contenido

Solución de problemas

loading

Tabla de contenido