Figure 6
Examples of flow rate settings
3 ml/hr
CHANGING THE FLOW RATE DURING AN INFUSION
1. Insert the rate-changing key into the dial of the Select-A-Flow* device.
2. Turn the dial until the new flow rate is selected. Make sure the selected flow rate setting
is aligned with the ml/hr
feel will allow the user to ensure selected flow rate is set.
3. Remove the key from the dial and put in a safe place for later.
DURING THE INFUSION:
•
A change in appearance and size of the pump may not be evident during the first 24
hours after start of infusion.
•
As medication is delivered, the pump will gradually become smaller.
•
Make sure:
•
Clamp is open
•
There are no kinks in the tubing
•
Filter vent is not taped or covered
•
Heat, ice or cold therapy is placed away from the flow controller
END OF INFUSION:
•
Infusion is complete when pump is no longer inflated.
•
Close clamp, disconnect and dispose of the pump according to your institution's
protocol.
Note: If pump did not perform as expected do not discard. Contact your representative for
product return instructions.
6 ml/hr
▼
mark on the face of the Select-A-Flow* device. The tactile
8
Figure 7
Incorrect alignment
Correct alignment