Leakage Clamp Meter
3.2.5 Circuit Continuity Test
Measuring range
Resolution
0.1Ω
- Overload protection: 600V DC or AC (RMS)
3.2.6 Temperature Test
Measuring range
Resolution
-20°C~0°C/-4°F~32°F
-0°C~400°C/32°F~752°F
1°C/1°F
400°C~1000°C/752°F~1832°F
- Overload protection: 600V DC or AC (RMS)
- The parameter does not contain thermocouple errors
3.2.7 Capacitance
Measuring range
Resolution
40.00nF
0.01nF
400.0nF
0.1nF
4.000μF
0.001μF
40.00μF
0.01μF
400.0μF
0.1μF
4.000mF
0.001mF
40.00mF
0.01mF
- Overload protection: 600V DC or AC (RMS)
- The parameter does not contain errors caused by base
capacitance and capacitance probe
Accuracy
If the resistance of circuit to
be measured is less than
40Ω, the meter's built-in
buzzer may sound.
Accuracy
±(
3.0
% reading +
5
digits)
±(
1.5
% reading +
5
digits)
±(
3.0
% reading +
5
digits)
Accuracy
±(
3.0
% reading +
8
digits)
35
Leakage Clamp Meter
3.2.8 Diode Test
Measuring range
Resolution
0.001V
- Forward DC current is about 1mA
- Backward DC voltage is about 3.2V
- Overload protection: 600V DC or AC (RMS)
4. Operating Guidance
4.1 Reading Hold Fuction
In the process of measurement, if reading hold is
required, press "LPF/HOLD" key, the value on the
display will be locked. press "LPF/HOLD" key again
to cancel reading hold state.
4.2 50Hz/60Hz Current Selection Fuction
In the process of current mode, press "LPF/HOLD" key
more than 2 sec, the meter will enter the LPF(50Hz/60Hz)
current measurement.
4.3 Maximum/Minimum Measurement Fuction
1) Press "MAX/MIN" key to enter MAX mode, the meter
will enter maximum measurement value; press
"MAX/MIN" key again, the meter will enter minimum
value measurement state; press "MAX/MIN" key
to repeat the above operations by recycling.
2) If the user presses "MAX/MIN" key more than 2 sec,
the meter will restore normal measuring range.
Function
Display approximate diode
forward voltage value