Materials :
Black PVC and AISI 316 Body
Special Glass Optics with oleophobic treatment
NBR and Silicon OR
Thread: 1" GAS
Measuring ranges: 0-1000 NTU, 0-4000 NTU
Measuring method: 90° Scattered light
Resolution:
-
0.01 NTU range 0-1000 NTU
-
0.01 NTU range 0-4000 NTU
Accuracy:
±2% at the measuring point range 0-1000 NTU
-
±5% at the measuring point range 0-4000 NTU
-
Ripeatability:
±5 NTU range 0-1000 NTU
-
±20 NTU range 0-4000 NTU
-
Responding time: T
Calibration: 1-point for scale with formazin standard solution
WorkingTemperature: 0÷50° C
Max Working Pressure: 4 bar
Mechanical Protection: IP68 Sensor+cable
Cable: 10m integral
Power Supply: 12...24Vdc
Max Absorption: 3W
Outputs: RS485 ( 4-20mA optional)
Cable colors coding
Order codes
9710597063
9710598063
9710599063
9720597063
9720599063
9730598063
9730599063
9740597063
9740599063
9760598063
S461TN Infrared Turbidity Probe – Data Sheet
TECHNICAL DATA
< 60s
90
RED
+24Vdc
BLACK
GROUND
YELLOW
A+ RS485
GREEN
B- RS 485
WHITE
+ 4-20 mA
BROWN
- 4-20 mA
S461TN AISI316 and PVC Turbidity probe 4-20mA output
S461TN AISI316 and PVC Turbidity probe RS485 output
S461TN/INS AISI316/PVC Insertion Turbidity probe 4-20mA output
S461TN PVC Turbidity probe 4-20mA output
S461TN/INS AISI316 Insertion Turbidity probe 4-20mA output
S461TN PVC Turbidity probe RS485 output
S461TN/INS AISI316 Insertion Turbidity probe RS485 output
S461TN AISI316 Turbidity probe 4-20mA output
S461TN/INS AISI316/PVC Insertion Turbidity probe RS485 output
S461TN AISI316 Turbidity probe RS485 output
S
4
6
1
T
N
I
n
f
r
S
4
6
1
T
N
I
n
f
r
T
u
r
b
i
d
i
t
y
p
T
u
r
b
i
d
i
t
y
p
DIMENSIONS
a
r
e
d
a
r
e
d
r
o
b
e
r
o
b
e
D
a
t
a
S
h
e
e
t
D
a
t
a
S
h
e
e
t
Page 19