Checklist; Electromagnetic Interference - Enraf Nonius Endolaser 422 Manual De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 122
6.3.1

Checklist

PROBLEM
Unit does not function
LCD display on front panel
does not light up
Some of the keys on the
front control panel do not
function correctly
The unit lights up but does
not emit energy
The unit functions as normal
but a relevant decrease in
efficiency of treatment is
evident.

7 Electromagnetic Interference

The Endolaser unit has been designed and manufactured according to current Directives regarding
Electromagnetic compatibility 2014/30/UE with the aim of providing adequate protection from harmful
interferences when installed in homes and health establishments.
All the necessary measures and tests were carried out at the Enraf-Nonius BV. Testing and Measure
Laboratory (LPMC) and in external specialised laboratories. The customer, upon prior request, may view
the reports relative to EMC measures within the company.
Customers have been given the possibility to verify reports relative to EMC measures within the company
upon request.
The working principles of the Endolaser unit for laser therapy do not generate significant radio-frequency
energy and are sufficiently immune to irradiated electromagnetic fields. No harmful interference is
therefore verifiable to radio electric communication, electro-medical monitoring equipment, diagnosis,
therapy and surgery, electronic devices in offices such as computers, printers, photocopiers, fax
machines, etc. or any other electrical or electronic equipment used in similar places as long as they are
ELECTROMAGNETIC COMPATIBLE.
EN
18
POSSIBLE CAUSE
Power cable incorrectly inserted into the
connector on the rear of the unit. Cable
worn out, damaged or interrupted.
The switch on the rear of the unit is
turned off. Fuses missing, blown or
interrupted.
Electronic control circuit has broken
down.
No power reaching the socket.
Presence of faulty components on
electronic control card
Faulty keys or buttons. Electronic
control circuit broken down
Parameters not set correctly. Laser
source does not function or is finished.
Faulty components on electronic control
circuit. Faulty supply on laser circuit.
Faulty or worn out laser source.
Possible break down in power generator
circuit of the unit.
SOLUTION
Insert the plug correctly into
the socket. Replace the worn
out or damaged power cable.
Turn on the switch. Replace
any missing, blown or
interrupted fuses.
Contact the local Enraf-Nonius
service centre.
Check the mains
Contact the local Enraf-Nonius
service centre
Contact the local Enraf-Nonius
service centre.
Check that the parameters
have been set correctly.
Check laser source emission
turned on.
Contact the local Enraf-Nonius
service centre.
Contact the local Enraf-Nonius
service centre.