10 Maintenance and troubleshooting
10.1 Cleaning
10.1.1 Cleaning of the apparatus
Switch off the unit and disconnect it from the mains supply.
The apparatus can be cleaned with a damp cloth. Use lukewarm water and a non-abrasive liquid
household cleaner (no abrasive, no alcohol content solution).
10.1.2 Cleaning of the radiators and HF-cable
The radiators can be disinfected using a cloth moistened with 70% alcohol. The radiators and HF-
cable may not be sterilized by expose to steam or gas. Clean the radiators after each use.
The radiators and HF-cable should be regularly inspected for damage and bad contacts. Damaged
radiators and a damaged HF-cable may not be used.
10.2 Fault conditions
10.2.1 Display's fail to light up
•
Check if the mains cable is connected to mains supply.
•
Check if the mains supply is available (power supply) for instance with a lamp.
If the lamp does not light up, check the mains fuses.
o
If the lamp lights up, check the fuses of the unit.
o
10.2.2 Error code Er XXX* on the display
The unit has discovered a fault using its "self-test". Switch the unit back on. If the code re-appears,
contact your supplier. The unit is probably defective.
*Er XXX: a number is mentioned on position X
10.2.3 Power cannot be set
•
HF-cable is not connected.
•
No radiator is connected.
•
No treatment time is set.
10.2.4 During treatment the HF-power switches off and a sound signal is audible (3 beeps
of 1 sec.)
•
The HF-cable is mistakenly disconnected from radiator or apparatus.
•
The HF-cable is damaged and must be replaced.
•
Bad contact occurred in the connection of the HF-cable to the apparatus and/or radiator.
10.3 Technical maintenance
We recommend to have the unit checked annually. This may be done by your supplier, or by another
agency, authorised by Enraf-Nonius. It is also recommended that a record of the service history is
kept for all activities relating to service and maintenance. In some countries this is even obligatory.
Maintenance and all repairs should only be carried out by an authorised agency. Enraf-Nonius will not
be held responsible for the results of maintenance or repairs by unauthorised persons.
Opening of the equipment by unauthorised agencies is not allowed and will terminate any claim to
warranty.
10.4 Expected Service Life
This device will remain suitable for its intended use as long as it is subjected to yearly maintenance by
a qualified service engineer as described in the service manual.
The service engineer will decide whether the device is suitable for use according to the specifications.