The RADARMED should only be operated with the original mains cable specified in the list of contents
delivered.
Operating the device with any other mains cable can lead to increased emissions or reduced interference
immunity of the device.
Guidance and manufacturer's declaration – electromagnetic interference
The RADARMED is intended for use in the electromagnetic environment specified below. The customer
or the user of the RADARMED should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
/ flicker emissions
IEC 61000-3-3
WARNING
The device should not be used when placed immediately next to or stacked on top of other devices. If
operation is necessary when immediately next to or stacked on top of other devices, the device should be
monitored to ensure it is operating as intended in this arrangement.
Guidance and manufacturer's declaration – electromagnetic immunity
The RADARMED is intended for use in the electromagnetic environment specified below. The customer
or the user of the RADARMED should assure that it is used in such an environment.
IMMUNITY
IEC 60601 test level
test
Electrostatic
± 6 kV contact
discharge
± 8 kV air
(ESD)
IEC 61000-4-2
Electrical fast
± 2 kV for power supply
transient/burst
lines
± 1 kV for input/output
IEC 61000-4-4
lines
Surge
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
IEC 61000-4-5
Voltage dips,
<5 % UT
short
(>95 % dip in UT) for 0,5
interruptions
cycle
and voltage
variations on
40 % UT
power supply
(60 % dip in UT) for 5
input lines
cycles
70 % UT
IEC 61000-4-
11
(30 % dip in UT) for 25
cycles
<5 % UT
(>95 % dip in UT) for 5 s
Power
3 A/m
frequency
(50/60 Hz)
magnetic field
081-400-095-49_EN
Compliance
Group 1
The RADARMED uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
Class B
The RADARMED is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low voltage
Class A
power supply network that supplies buildings used for
domestic purposes.
Complies
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % UT
(>95 % dip in UT) for 0,5
cycle
40 % UT
(60 % dip in UT) for 5
cycles
70 % UT
(30 % dip in UT) for 25
cycles
<5 % UT
(>95 % dip in UT) for 5 s
3 A/m
Electromagnetic environment – guidance
Electromagnetic environment
Floors should be wood,
concrete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the RADARMED requires
continued operation during
power mains interruptions, it is
recommended that the
RADARMED be powered from
an uninterruptible power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.