Technical Data
Current-/Power consumption: ........max. 10mA/ 0,1W
Internal Fuse: ..................................F 630mA / 250V 5x20mm
Measurements/hour: ......................100
Dimensions (LxBxD): ......................235 x 105 x 70 mm
Weight: ............................................approx. 700 g (incl. battery)
Environment Conditions:
Temperature range: ........................0 °C...40 °C
Humidity: ........................................10°C – 30°C at 80%
Altitude: ..........................................up to 2000 m
Protective insulation. ......................IP 50
Protection Class ..............................II
Overvoltage Categ.:..........................CAT III / 300V against ground
Pollution Degree: ............................2
Nominal error current
Pretest-current
I∆N
10mA
I∆N
30mA
I∆N
100mA
I∆N
300mA
I∆N
500mA
I∆N
1000mA
I∆N
100mA selective I∆N
300mA selective I∆N
500mA selective I∆N
24 month warranty
UNITEST instruments are subject to strict quality control. However, should the instrument function improperly dur-
ing normal use, you are protected by our 24 month warranty (valid only with invoice).
We will repair free of charge any defects in workmanship or material, provided the instrument is returned unopened
and untampered with.
Damage due to dropping or incorrect handling are not covered by the warranty.
If the instrument shows failure following expiration of the warranty, our service department can offer you a quick and
economical repair.
Subject to technical changes without notice !
24
Test current I∆N
10mA
3,3
30mA
10
100mA
33
300mA
100
500mA
167
1000mA
333
200mA
3,3
600mA
100
600mA
167
Test time
<300 ms
<300 ms
<300ms
<300 ms
<300 ms
<300 ms
<300 ms
<300 ms
<300 ms