CARE & SAFETY INFORMATION
•
To verify the calibration of this device, please contact the manufacturer.
•
This device may be used only for the purpose described in this booklet. The
manufacturer cannot be held liable for damage caused by incorrect application
and use.ITROD
•
The service life of the cuff may vary by the frequency of washing, skin condition
of the user, and adherence to proper storage conditions. The typical service life
is 10000 measurements.
•
It is recommended that the performance should be checked every 2 years and
after maintenance and repair, by retesting at least the requirements in limits
of the error of the cuff pressure indication and air leakage (testing at least at
50mmHg and 200mmHg).
•
The device doesn't need to be calibrated within two years of reliable service.
•
Keep the unit out of reach of infants, young children or pets to avoid inhalation
or swallowing of small parts. It is dangerous or even fatal.
OPERATING CAUTIONS AND WARNINGS
•
Prolonged over-inflation may cause congestion, swelling or bruising in some
people.
•
Wait a minimum of 2 minutes between measurements (20-minutes is
recommended); excessive frequent measurements may restrict blood circulation.
•
Do not use this device while connected to any monitoring equipment on the same
limb.
•
Consult with your healthcare professional for guidance for use if you have any of
the following: intravascular shunt; mastectomy on the arm corresponding to the
measurement side of the body.
•
Consult with your healthcare professional before using the device in pregnancy
including pre-eclampsia, or if diagnosed with arrhythmia or arteriosclerosis.
•
If you experience any discomfort or an abnormality occurs, stop using the
device immediately; to stop the device during operation, press the U1 or U2
(POWER) buttons and the cuff will automatically deflate.
ENGLISH • 5