Installation in PSS8 Probeholder
Install the probe in the pss8 respecting the position shown in the figure
above.
To facilitate the correct insertion of the probe body there is a signaling tag
that indicates the direction of the flow.
When the probe is installed in PSS8 probeholder, the calibration curve must
be changed to compensate the offset of the wall of the probeholder itself,
which is why when ordering you must specify if the installation will be
performed in bypass with the PSS8 probeholder.
CAUTION!!! each probe holder PSS8 has an inherent structure that may
slightly interfere with the reading, so each system sensor + probeholder
must be calibrated together.
it is recommended to install a tap at the water output of the probeholder in
order to reduce the flow and thus create a pressure inside of the probeholder
itself able to favor the elimination of any air bubbles.