Diamond Diagnostics RD-945375D Instrucciones De Procedimiento

Electrodo de pco2
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 10

Enlaces rápidos

Manufacturer and Product Information
Diamond Diagnostics, 333 Fiske Street, Holliston, MA.
For Technical Assistance call:
Diamond Diagnostics Technical Services at 1-508-429-0450
Contents:
Intended Use:
Storage:
Precautions:
Procedure
Installation:
1.
Unpack the new pCO
2.
Add 5 drops of Electrode cleaner, RD-943905E, into the Cuvette provided with the electrode.
3.
Soak the electrode in the vial for 1 hour.
4.
Remove the electrode and rinse with plenty of distilled or deionized water.
5.
Empty the Cuvette and rinse with distilled or deionized water. Fill with distilled/deionized water up to one third the height of the Cuvette.
6.
Soak the electrode in distilled/deionized water for 24 hours. The electrode may drift excessively if soaked for a shorter time.
7.
Remove the cover of the temperature controlled electrode unit.
8.
Press the pCO
amplifier housing down, pull the metal tab outward, and lift the amplifier housing.
2
9.
Remove the old pCO
10. Remove the newly prepared electrode from the Cuvette with water. Gently shake excess water from the electrode.
11. Obtain a box of pCO
12. Press the electrode firmly into the jacket.
13. Remove the jacketed electrode from the box.
14. Look for bubbles between the membrane and the electrode. If bubbles are observed, gently move the electrode up and down in the
electrolyte to release the air bubbles. Press the electrode back into the jacket again.
15. Using a dry clean cloth or paper towel, thoroughly wipe and dry the gold contacts of the electrode.
16. Place the newly prepared electrode in the correct section of the measuring chamber.
17. Gently press the amplifier housing down onto the electrode.
18. Record in the maintenance schedule that the pCO
19. If all maintenance steps have been performed, place the cover over the temperature controlled electrode unit and press Restart
Analyzer key below the display in scheduled maintenance menu
Warranty
Warranty = 12 months:
The names of various manufacturers, their instruments, and their products referred to herein may be protected by trademark or other law, and are used herein solely for purpose
of reference. Diamond Diagnostics expressly disclaims any affiliation with them or sponsorship by them.
1 – pCO
Electrode
2
1 – Cleaning Solution (RD-943905E)
1 – Cuvette
Use with Radiometer ABL 5xx Series analyzers.
Store the pCO
electrode at room temperature (18-25˚C) until you are ready to use it in your system.
2
Following pCO
electrode replacement or membrane change, automatic calibrations are changed to the following
2
Sequence:
1.
First 4 hours, alternate in 30 minute intervals Gas Cal 1 and Gas Cal 2, ending in Gas Cal 2.
2.
Next 20 hours, every hour Gas Cal 1, every second hour alternate with Gas Cal 2.
electrode. Preparation requires 25 hours.
2
electrode.
2
membranes in jackets. Remove the foil cover.
2
2
Warranty is calculated from date of shipment from Holliston, MA.
electrode has been replaced.
ECO# 6195 SOP05-1096F Revision 07 Effective Date: 10/23/12
pCO
Electrode
2
RD-945375D
Page 1 of 10
loading

Resumen de contenidos para Diamond Diagnostics RD-945375D

  • Página 1 The names of various manufacturers, their instruments, and their products referred to herein may be protected by trademark or other law, and are used herein solely for purpose of reference. Diamond Diagnostics expressly disclaims any affiliation with them or sponsorship by them.
  • Página 2 Electrode RD-945375D Bulgarian Diamond Diagnostics, 333 Fiske Street, Holliston, MA. Diamond Diagnostics Technical Services at 1-508-429-0450 1 – pCO 1 – (RD-943905E) 1 – ABL 5xx Series (18-25˚C) Gas Cal 1 Gas Cal 2, Gas Cal 2. Gas Cal 1, Gas Cal 2.
  • Página 3 Elektroda RD-945375D Czech Informace o produktu a výrobci Diamond Diagnostics, 333 Fiske Street, Holliston, MA. Technická asistence na čísle: Diamond Diagnostics technický servis 1-508-429-0450 Obsah: 1 – pCO Elektroda 1 – Čistící roztok (RD-943905E) 1 – Kyveta Doporučené použití: Kompatibilní pouze se systémy Radiometer ABL série 5xx.
  • Página 4 Le nom de certains producteurs ainsi que leurs produits mentionnés dans ce document peuvent avoir une marque déposée protégée par le Droit des marques, ils sont donc utilisés uniquement pour informations et références. Diamond Diagnostics refuse toutes sortes de relations privilégiées avec ces producteurs ou leurs produits.
  • Página 5 Die verschiedenen Namen und Hersteller von Geräte oder Produkte,die in diesem Dokument erwähnt wurden,sind gesetzlich geschützt,in einigen Fällen auch eingetragene Warenzeichen und Informationen dienen ausschliesslich in diesem Dokument geschützt.Diamond Diagnostics fest weigern,mit diesen Herstellern in keiner weise oder durch die Bereitstellung von Unterstützung in Verbindung gebracht werden.
  • Página 6 A különböző gyártók neve és gépeik, illetve termékeik, amelyekről ezen dokumentum említést tesz, törvény által védettek, vagy bejegyzett márkák, és kizárólag tájékoztatási céllal kerültek bele ezen dokumentumba. Diamond Diagnostics határozottan visszautasítja azt, hogy bármilyen módon összefüggésbe hozzák ezen gyártók üzleti tevékenységével, illetve annak feltételezését, hogy ezen gyártók bármilyen módon támogatnák a Diamond Diagnostics-ot.
  • Página 7 I nomi , gli strumenti e i prodotti dei vari produttori i quali sono stati nominati in questo documento sono protetti dalla legge, a volte sono marchi depositati e si possono trovare in questo documento soltanto per dare delle informazioni.Diamond Diagnostics rifiuta di essere relazionato con questi produttori o di essere sovvenzionati da loro.
  • Página 8: Garantia

    Os nomes de vários fabricantes, os seus instrumentos, e os seus produtos mencionados neste lugar podem ser protegidos pela marca de comércio ou outra lei, e são usados neste lugar sozinho para o objetivo da referência. Diamond Diagnostics expressamente nega qualquer associação com eles ou patrocínio por eles.
  • Página 9 Nume de diferite producători, instrumentele şi produsele lor prevăzute în acest document ar putea fii protejate de mărci comerciale sau de lege şi sunt folosite numai pentru referinţă. Diamond Diagnostics refuză formal orice legătură cu producătorii şi mărcile menţionate în acest document, sau orice contribuţie cu acesti producători.
  • Página 10 Los nombres de varios fabricantes, sus instrumentos, y sus productos mandados a aquí pueden ser protegidos por la marca registrada u otra ley, y son usados aquí únicamente para el objetivo de la referencia. Diamond Diagnostics expresamente rechaza cualquier afiliación con ellos o patrocinio por ellos.