Transcend 3 miniCPAP Guía Rápida página 11

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 60
Transcend 3 miniCPAP Quick Guide
Note:
In the event of power loss during ramp, the Transcend 3 miniCPAP will
resume at the full prescribed pressure as soon as power is restored.
Using the EZEX Function
The EZEX function is a special feature that decreases therapy pressure on exhalation. This
feature is designed to provide additional comfort to the patient by reducing the amount of
resistance they experience as they exhale.
There are four EZEX settings: OFF, 1, 2 or 3; progressively increasing the amount of
pressure relief from none to maximum. The software section of the user manual displays
how to modify EZEX settings.
Note: To download the Transcend 3 miniCPAP user manual visit
www.MyTranscend.com/support or call Somnetics customer service at 877-621-9626 to
request a copy.
Ending Therapy
To end the delivery of therapy while the blower is on, press the Power button to
deactivate the blower and return the device to Standby Mode. It is recommended the user
initiate the Drying Mode function after each therapy session to dry the device interior.
Drying Mode
At the end of each therapy session it is recommended the user initiates the Drying Mode.
To initiate Drying Mode depress the Ramp button and press the Power button
simultaneously. When in Drying Mode the blower runs at a low speed for 30 minutes.
During Drying Mode the LED lights remain off and blower pressure is not regulated to
provide therapy. Using the Drying Mode flushes air through the system to remove traces
of moisture from the interior of the device and airway circuit. After the 30-minute drying
cycle, the blower will turn off and the device will automatically enter Standby Mode.
Page 11
Tabla de contenido
loading

Tabla de contenido