Performing A Control Solution Test - Fora V20 Manual Del Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Performing a Control Solution Test

Step 1
Insert the test strip to turn on the meter.
Insert the test strip into the meter. Wait for the meter to display the
test strip and blood drop symbols.
Step 2
Press
to mark this test as a control solution test.
With " CTL " displayed, the meter will not store your test result in
memory. If you press
is no longer a control solution test.
When doing the control solution test, you have to mark it so that the test
result will not be stored in the memory. Failure to do so will mix up the blood
glucose test results with the control solution test results in memory.
again, the " CTL " will disappear and this test
WARNING
19

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

Td-4245

Tabla de contenido