INSTRUCTION MANUAL 80200 Rev B February 2014
J
.P. SELECTA s.a.u.
(this document can be modifi ed without notice)
6.5 Calibration
The CD-2005 can be calibrated in one point. It is necessary to use a calibration pat-
tern near the reading point.
Calibration of the CD-2005:
•
Having conductivity pattern with a determined value (P)
•
Check that the conductivity constant of the probe is correct: 0.01, 0.1, 1 or 10
•
Introduce the conductivity cell inside the pattern liquid
•
Wait that the reading will be stable
•
Make note of the reading (L)
•
Calculate the correction factor: C = P / L
•
Press one of the keys «ADJ» (10), the display will show the current value
•
With the keys «
•
Press «ENTER» (12) to validate the new correction value
The conductivity meter is calibrated.
Note: Usually the conductivity meter must be calibrated:
•
On starting up
•
When the probe is changed
6.6 Conductivity cell handling
•
Follow the own intructions from the conductivity cell if it has.
•
Avoid to hit the cell.
•
Clean with water after every measurement.
7 Warranty
The warranty covers the defective components or faulty assembling of the unit for 1
year since the purchasing date.
Keep the purchasing invoice as the only valid document for the purchasing date.
The warranty does not covers the damages caused for a unit malfunction. The unit
must be suitably qualifi ed before to be used for valuable samples processing.
Return good are not accepted without previous J.P. SELECTA, s.a.u. approval.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
» and «
» enter the calculated value "C"
Tel 34 937 700 877 Fax 34 937 702 362
Page: 11