Changing Battery
Replace battery when Low or Dead Battery
Symbols appear in Display or Meter does
not turn on.
1. Lift tab on Battery Cover.
2. Turn Meter over, tap gently
to loosen and remove battery.
3. Insert new battery, positive ("+")
side up. Close cover.
Note:
Use non-rechargeable 3V
lithium battery (#CR2032).
4. Discard old battery in appropriate container.
5. Turn Meter on. If Meter will not turn on, check that
battery was installed properly. If not, reinsert battery and
try again. Call for assistance if problem persists.
Warning! Battery is not rechargeable. If you have a cable or
a cradle for downloading results to a computer, DO NOT plug
the usb cable end into an electrical outlet. Trying to recharge
the battery or power the meter by plugging into an electrical
outlet will cause meter to catch on fire or melt.
Caution!
Note:
Replacing battery may affect date and time settings.
Check date and time by going to Set Up and scrolling
through the time/date settings by pressing
Battery is not rechargeable. Battery might
explode if mishandled or incorrectly replaced. Do
not dispose of battery in fire. Do not disassemble
or attempt to recharge battery. Dispose according
to local/country specific regulations.
33
Low
Dead
.