Step [ 2 ] : Prepare Bresle Method Patch
Option A: DeFelsko PosiPatch
a) Turn the PosiPatch and magnetic ring upside down. Place
the PosiPatch into the magnetic ring ensuring the arrows align
with the injection port. Do not touch the PosiPatch's interior.
b) Carefully place the magnetic ring on the steel surface.
When used on vertical surfaces, position the port between the
2 o'clock and 10 o'clock points before attaching the ring to the
surface.
c) Slowly inject 3.0ml of DI water into the port using the plastic
dispensing tip. The air within the PosiPatch automatically
escapes through the air-permeable membrane
a
(
)
When used on horizontal surfaces, fully insert the dispensing tip
into the PosiPatch. On vertical surfaces, insert the dispensing tip
only as far as is needed to see the tip inside the PosiPatch.
Option B: Bresle Method DeFelsko Patch (or equivalent)
a) Remove the backing and foam center from the patch, taking
care not to touch the interior surface.
b) Press the patch firmly onto the surface making sure most of
the air escapes.
c) Pierce the top of the patch at the foam perimeter, allowing
the needle to penetrate into the center of the patch.
b
(
)
4
c
(
)