General Description; Technical Features - Seneca K109TC Manual Del Usuario

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

K109TC
EN
TEMPERATURE-INSULATED CONVERTER FOR THERMOCOUPLES

General Description

The K109TC instrument is a 4-point galvanic insulator. It converts a temperature signal
read by a thermocouple into a signal normalised in voltage or current; besides it has a
passive output which commutates at the overcoming of a threshold set through a button on
the front.
The module's main features are its compact size (6.2 mm), attachment to a 35 mm DIN
rail, bus-connector power supply option, quick connection by spring terminals, and easy
configuration in the field by DIP-switch.

Technical Features

Power supply :
Consumption :
Input :
Tables :
Measurement Range :
Minimum span :
Impedance :
Test Current :
(1)
CMRR :
(1) (2)
DMRR
:
(1)
The values are valid at the set rejection frequency, with filter ON.
(2)
For disturbance values such as the input signal peak does not exceed the limit of
acceptability.
Thermocouple
J
K
E
N
S
R
B
T
(*): Up to 250 °C, the output is considered equivalent to a null temperature.
19,2..30 Vdc
max 24 mA at 24 Vdc
Thermocouples types: J, K, E, N, S, R, B, T
EN60584-1 (ITS-90)
Depending on the thermocouple type (see Range and
Precision of the Input Table), on the dip-switches
selection (see
100 °C
10 MW
<50 nA
>135 dB, refered to power-supply side
>40 dB
Range and Precision of the Input
Admitted Range
-210..1200 °C
-200..1372 °C
-200..1000 °C
-200..1300 °C
-50..1768 °C
-50..1768 °C
250..1820 (*)°C
-200..400 °C
MI00111 -E
Setting of the Dip-Switches
Mean Error
0,025 % + 0,29 °C
0,025 % + 0,4 °C
0,025 % + 0,2 °C
0,025 % + 0,42 °C
0,025 % + 1,34 °C
0,025 % + 1,19 °C
0,025 % + 1,87 °C
0,025 % + 0,31 °C
3
section)
Resolution
0,12 °C
0,17 °C
0,92 °C
0,19 °C
0,66 °C
0,59 °C
0,9 °C
0,13 °C
ENGLISH - 1/8