Operational Diagram; Troubleshooting - OK International DX-300 Guia Del Usuario

Dosificadores de precisión
Idiomas disponibles
  • ES

Idiomas disponibles

7
OPERATION DIAGRAM
MODE
:Press then release
Prog
Time Counter
:Press and hold for approximately 4 seconds then release
POWER
MODE
SET
Normal Operation
P1
OFF
P1
CONTINUE
P1
2.250
00004
*************************
Change Mode
Power ON/OFF
P1
SETUP MODE
SETUP MODE
P1
2.250
00004
P1
TIME SETTING
SETUP MODE
SETUP MODE
P1
2 .250
2
00004
P1
CONTINUE
Save and Exit...
P1
2.251
00004
COUNTER RESET
SETUP MODE
Save and Exit...
P1
2 .251
2
00004
:Press and hold for approximately 2 seconds then release
CONTINUE
2.250
00004
INTERRUPT
CONTINUE
2.250
00004
P
P1
2.250
00004
PURGE
CONTINUE
P
2
2.250
00004
CONTINUE
P
1
2.250
00004
CONTINUE
P
1
2.250
00004
Select Program
Setup
CONTINUE
Time setting
2.250
00004
in CONTINUE mode
SETUP MODE
2 .250
2
00004
P1
2
2 .250
00004
SETUP MODE
2 .251
2
00004
P1
2 .250
2
00004
SETUP MODE
P1
2 .251
2
00004
CONTINUE
P1
2.251
00000
7
8

TROUBLESHOOTING

PROBLEM
POSSIBLE CAUSE
LCD does not
No power input
light
System will not
Foot switch not
actuate
plugged in or
improperly
plugged in
Defective foot
switch
Broken wire or
loose connection
inside unit
Defective
solenoid
Defective PC
board
System will not
Insufficient air
pressurize
pressure
Air hoses not
plugged in
Regulator
defective
System will not
Vacuum setting is
pull vacuum
too low
Defective
solenoid
Defective
vacuum venturi
Defective
vacuum needle
valve
Inconsistent
Air bubbles in
dispensing
adhesive
Dispense time is
too low
Dispense needle
started to clog
CORRECTION
Check power cord
connections
Turn on power
Check foot switch
connection
Foot switch needs
to be repaired or
replaced
Unplug power cord
and disconnect air
supply. Remove
cover and check
for broken wires or
loose connections
Replace solenoid
Replace PC board
Increase air supply
pressure
Check connection
Replace regulator
Increase vacuum
setting
Replace solenoid
valve
Replace venturi
Replace needle
valve
Reduce vacuum
setting
Increase
dispensing time
Replace needle
8
loading