RS Pro RS-172 Manual De Instrucciones página 5

Registrador de datos de temperatura y humedad
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

Temperature & Humidity Datalogger/English
Datalogger Setup
• Click on the icon
on the menu bar. The Setup window will appear as shown below
Descriptions for each field in the Setup window are listed directly below the illustration:
• The Samprate Setup field instructs the DATA LOGGER to log readings at a specific rate.
You can input specific data at the left Edit box and select the time unit at the right Combo
box.
• The LED Flash Cycle Setup field can be set by the user(Select "No Light",It will not
flash,and the battery life is the longest.)
• The Manual and Automatic select buttons allow the user to start data logging
immediately when the Setup window is exited (Automatic) or at a later time (Manual).
• The Alarm Setup portion allows the user to set HIGH and LOW temperature, relative
humidity limits and disable the flashing of the DATA LOGGER's LEDs thereby conserving
battery energy.
• The Circulating Record and No Circulating select buttons instruct the DATA LOGGER
continue logging data to cover the earlier data(Circulating Record) or stop logging(No
circulating) when the DATA LOGGER is full.
Click on the SETUP button to save changes. Press the DEFAULT button to set the Logger
to factory default condition. Press the CANCEL button to abort the setup.
Notes: Any stored data will be permanently erased when Setup is finished. To enable you
to save this data before it is lost, click Cancel and then you need to download data.
The battery will in all likelihood run out before logger will have finished specified sample
points. Always ensure that the remaining charge in the battery is sufficient to last the
complete duration of your logging exercise. If in doubt, we recommend that you always
install a fresh battery before logging critical data.
6
Datalogger Download
To transfer the readings stored in the Logger to the PC:
A.Connect the DATA LOGGER to the same USB port used when the DATALOGGER was
initialized.
B.Open the HT Datalogger software program if it is still not running
C.Click the Download icon
D.The window shown below will appear. Press DOWNLOAD to begin transferring readings
If readings are successfully transferred, the Data graph window will appear.
The top field shows detailed infomation of the readings(Time, Sampling rate, data
numbers,Alarm Hi,Alarm Low,Max data,Min data and Avg data),