NorthEast Monitoring OxyHolter DR181 Guía De Conexiones

Ocultar thumbs Ver también para OxyHolter DR181:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 7

Enlaces rápidos

5-Electrode, 2-Channel for OxyHolter
Patient Preparation
Select area over bone, avoiding muscle areas and breast tissue.
Clean electrode sites aggressively with alcohol and use a razor to
remove hair.
Abrade electrode sites with a scrub pad or gauze. Allow sites to
dry before applying electrodes.
Firmly snap electrodes to the lead wires.
Apply electrodes to appropriate sites. Press adhesive border
firmly for consistent adhesion.
To help alleviate tension, loop and tape lead wires down.
Tape
Tape
DR181 Hook-Up Guide for OxyHolter
Electrodes
--
+
Ch 1
mod. v5
WHITE
RED
Ch 2
mod. V1
BLACK BROWN
GREEN (Ground)
Electrode
Electrode
How to attach the probe and finger wrap
1. Remove the paper backing from the finger wrap
2. Attach the oximetry probe with cable to its sticky side, lining up the two bumps on
the probe with the cutouts in the wrap
3. Position the patient's finger on the probe. Be sure to line up the tip of the finger
with the center line on the wrap
4. Attach the wraps to the patient's finger in the following order:
a. Attach the two small side wings
b. Fold the wrap and probe back over the fingernail
c. Wrap the third small wing
d. Wrap the long wing around the finger
5. Use tape around the base of the patient's finger to secure the cable to the finger.
Apply additional tape to hand to secure cable.
Step 2
Step 1
NorthEast Monitoring, Inc.
141 Parker Street, Suite 111
Maynard, MA 01754 USA
advancing Holter technology
NEMM034 - Rev-C - English - January 2016
Finger Wrap
Step 4a
Step 3
Step 5
MediMark Eruope, SARL
11 rue Emile Zola
BP 2332
F-38033 Grenoble CEDEX 2
France
phone: [+1] 978-461-3992
fax: [+1] 978-461-5991
www.nemon.com
toll free USA: 866-346-5837
0143
Tabla de contenido
loading

Resumen de contenidos para NorthEast Monitoring OxyHolter DR181

  • Página 7: Guía De Conexiones Para Oxyholter Del Dr181

    BP 2332 • Haga un doblez en los alambres y fíjelos con cinta adhesiva para F-38033 Grenoble CEDEX 2 0143 reducir la tensión. France NorthEast Monitoring, Inc. Cinta Adhesiva Tape Electrodo Electrode 141 Parker Street, Suite 111 phone: [+1] 978-461-3992...
  • Página 8: Guía De Comienzo Rápido Para Oxyholter Del Dr181

    La pantalla presentará la fecha, la hora, la potencia de las pilas, “DR181”, la versión del software y “NorthEast Monitoring”. Pulse Próx para acceder al menú principal. 3. Si lo desea, puede ajustar los valores pulsando el botón en el teclado. En Valores, pulse Más para presentar opciones de valores adicionales.

Tabla de contenido