7-1 Cleaning
1) Remove dirt and debris from the product. Do not use a wire brush.
2) Wipe clean with alcohol-immersed cotton swab or cloth.
CAUTION
• To clean the product never use any solvent such as benzine or
thinner.
7-2 Lubrication
NSK PANA SPRAY Plus
Apply NSK PANA SPRAY Plus every time after each use and/or before
autoclaving.
1) Remove the motor from the hose.
2) Mount the arrow-head spray nozzle tip into the spray can port.
(Fig. 6)
3) Insert the Spray Nozzle into the Drive Air Port of the motor. Hold the
motor and spray for approximately 2-3 seconds. Apply lubricant
until it expels from the exhaust air part for at least 2 seconds.
(Fig. 7)
4
CAUTION
• W hen applying spray be sure to hold the motor firmly to prevent
the motor from slipping out of the hand due to the spray pressure.
• H old the spray can upright.
Lubrication Oil
1) Remove the motor from the hose.
2) Apply 1-2 drops of lubrication oil into the Drive Air Port shown in
the Fig. 7.
3) Connect the motor to the hose. Operate the motor for about 10
seconds.
7-3 Sterilization
Sterilize the product by autoclave sterilization. Remove the handpiece
after each patient and sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or
3 min. at 134°C.