English
INTRODUCTION
The TRITAN™ 365 AC/DC series UV LED Inspection lamps are equipped with a cord ending in a male
connector plug and an AC/DC adapter cord appropriate for the country of destination. The kit includes
a 12V 3000mAHr NiMH rechargeable battery in a case with an adjustable shoulder strap. Each smart
battery charger is equipped with the cord appropriate for the country of destination (120V, 230V, 240V or
100V). All components are packed in a sturdy, padded carrying case.
Ultraviolet radiation emitted from this product. Avoid exposure. ALWAYS WEAR PROTECTIVE
CLOTHING. EXPOSURE MAY CAUSE PREMATURE AGING OF THE SKIN AND CANCER.
ALWAYS WEAR PROTECTIVE EYEWEAR; FAILURE TO DO SO MAY RESULT IN SEVERE
BURNS OR LONG TERM INJURY TO EYE. Never look directly into the lamp. Exposure can
cause eye and skin allergy and allergic reactions. Medications or cosmetics may increase
your sensitivity to ultraviolet radiation. Consult physician before operating this product if
you are using medications or have a history of skin problems or believe yourself especially
sensitive to sunlight.
IMPORTANT SAFETY INSTRUCTIONS
• Read all instructions.
• Do not operate the lamp with a damaged cord or if the lamp has been damaged.
• Never use this equipment in any manner not specified in these instructions because the protection
provided by the equipment may be impaired.
• The TRITAN 365 is not approved for use in hazardous atmospheres. Do not attempt to use it in areas
requiring explosion-proof lighting.
CHARGING
The BP-30 battery pack must be fully charged before first use.
• Check that the voltage characteristics of the battery charger match the voltage of the
AC outlet.
• Align the male connector of the smart charger (BR-150A) to the female connector of the battery pack
(BP-30) and plug them together. Twist the ring on the charger connector to make a tight connection.
• Attach the AC cord supplied to the BR-150A charger and plug the unit into a properly regulated AC outlet.
• The indicator light on the charger will show Red while it is charging and Green when the battery is
charged. A full charge will take approximately 2 hours.
• After the battery is charged, first unplug the charger from the outlet, then unplug it from the battery pack.
WARNING: Use only the charger supplied to recharge the battery. Other chargers may harm the battery
and will void the battery's warranty.
For AC Operation
• The connector on the end of the 3.5-ft (1.06 m) cord on the lamp handle has an alignment peg, and
the corresponding connector on the AC power cord has a groove. Match the groove to the peg and
carefully screw the associated connectors together. Attach the AC adapter plug to the corresponding
AC power source that complies with the electrical requirements specified on the label.
• The lamp has a three-position rocker power switch (II–UV; O–off; I–white light) on the handle. The fan
turns on automatically and the light above the switch illuminates green when the UV LEDs are in use.
For DC Operation
• The connector on the end of the cord on the lamp handle has an alignment peg and the
corresponding connector on the fully charged BP-30 battery pack has a groove. Match the peg to the
groove and carefully push the associated connectors together.
• The lamp will operate continuously for approximately 3.5 hours on the standard, lower-intensity
setting, and for approximately 2.5 hours on the higher-intensity setting. When the indicator light on the
back of the handle turns Red the battery pack must be recharged.
You must wear the UV-absorbing spectacles supplied. The lamp should be used in
UV-A/White Light LED NDT Inspection Lamps
Part Nos. TRI-365MHA, TRI-365MDA, TRI-365MDBA
FOR PROFESSIONAL EXAMINATION USE ONLY
low-light conditions to achieve the best inspection results.
TRITAN™ 365 AC/DC SERIES
DANGER
IMPORTANT
2