HEINE NT300 Manual De Instrucciones página 12

Ocultar thumbs Ver también para NT300:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 24
The NT300 may only be used at the mains voltage specified on the identification label. Unplug the unit
from the mains if not in use over long periods.
In the event of fluid ingress in the housing, the charger should be taken out of service and checked by an
authorised person.
The NT300 must not be used in rooms where an explosion hazard is present.
Follow the instructions for the charger and rechargeable handles.
The following handles are not compatible with the NT300 charger:
BETA TR, BETA R, BETA cord handle, mini3000 battery handle.
Plug the mains transformer into the mains and the NT300 is ready for use.
NT300 ensures that batteries are kept in good condition irrespective of handle size.
When replacing the handles in the charging compartments, ensure that the internal contacts in the
compartment are clean and that the instruments are switched off.
Instruments placed next to each other must not touch each other.
If different-sized handles are used, it is possible to fit adaptors so that each handle is held in place correctly.
Reduction rings for:
BETA Handle
F.O. angled handle
Standard F.O.
Short F.O. laryngoscope handle
Ideally, the rechargeable handle should always be replaced and stored in the charger. Handles cannot be
overcharged.
The rechargeable battery can only be charged when fitted to the handle.
The NT 300 must be disconnected from the mains supply when the adaptors are being fitted.
If the second charging compartment is not in use it can be sealed off by the dust cap which is supplied with
the unit.
The pulsating light around the charging compartment shows that current is flowing and that the battery is being
recharged.
When the light is permanently lit, the charging sequence is completed. The battery in the handle is fully-charged
and is ready to supply maximum power.
If the light is switched off, this indicates that the charging compartment is empty or the unit is defective.
Hygienic Cleaning / Processing:
The device must be cleaned by sufficiently qualified personnel.
We recommend that external surfaces of the NT300 be wiped clean with a disinfectant cleaning wipe.
The external surfaces may also be wiped clean with a cloth moistened with water or with a neutral pH or slightly
alkaline cleansing agent.
Any disinfectant used must be approved by the disinfectant manufacturer for use on plastics.
The manufacturer's instructions must be followed.
Spray disinfection or immersion disinfection is not permitted since there is a risk of moisture ingress,
which would damage the device.
Disinfection:
The article can be neither machine-cleaned, nor thermal disinfected or sterilized.
These processes lead to irreparable damage to the product.