MAINTENANCE
The NEW MAMILAT breast pump does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use. Unpack the instrument and
always check integrity of plastic parts and feeding cable, they might have been damaged during
previous use. Connect cable to electrical network and turn switch on.
Close the aspirator outlet with your finger and with suction regulator in maximum vacuum position
check that the vacuum power.
Rotate the knob from right to left and check the aspiration regulating control.
The vacuum power should go down to the minimum value.
Verify that loud noises are not present, these can indicate wrong functioning.
A protection fuse (F 1.6 A 250V and F 4 A 250 V for voltage 110 V / 60 Hz) not reachable from
exterior protects the instrument.
For fuse replacing please contact GIMA S.p.A. technical assistance.
Fault type
Cause
Solution
1. No aspiration
Bottle cap badly screwed
Unscrew the cap, then rescrew it correctly
down
2. No aspiration
Cap seal not in its seat
Unscrew the cap and insert the seal
properly in its seal
3. No aspiration due to
Filter blocked
Replace filter
outflowing of milk from
overflow bottle
4. The motor doesn't
Activation of the thermal
Wait for some times to avoid the
work (green light on the
protection on the motor
deactivation of the thermal protection,
power switch on)
after that the motor start again to work
Faults 1 - 2 - 3 - 4
None of the remedies has
Contact the seller or GIMA S.p.A. After-
achieved the desired
sales Assistance Service
results
If the overfill security system it's activated don't proceed with the liquid aspiration.
The bacteriological filter will be stopped the aspiration.
If both the security system doesn't work, there is the possibility that the liquid comes inside the device,
in this case return the device to CA-MI technical service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT GIMA S.p.A. TECHNICAL SERVICE.
GIMA S.p.A. DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL
SERVICE CHECKING, APPEARS TO BE TAMPERED.