R e f r i g e r a t e d B a t h s U S E R ' S M A N U A L
siting Considerations
MINIMUM
150 mm (6")
setting up the unit
P A G E 10
All the units are air cooled, air being pulled in through the front
and expelled from the rear. The ventilation grilles should never
be obstructed, therefore, as this will detract from the performance.
Also the fridge units should never be situated in an enclosed space
or near an extractor fan, this too will detract from the performance.
1. check the voltage marked on the unit's rating plate against that
of the mains supply. connect the mains cable to a suitable plug
as in the following table. note that the unit must be earthed
to ensure proper electrical safety.
Connections
230V
Live
brown
Neutral
blue
Earth
yellow and green green
The fused plug supplied on the mains cable for use in the UK
with the RB-5A is fitted with the following fuse to protect the
cable.
Supply
fuse fitted
240V
5 Amp
For the RB-12A and RB-22A, if a fused plug is fitted then a fuse
of the following value must be fitted
Supply
fuse fitted
240V
13 Amp
The fuse in the unit protects the unit and the operator.
Note that units marked 230V on the rating plate work equally
well at 220V; units marked 120V on the rating plate work
equally well at 110V.
MINIMUM
150 mm (6")
110/120V
black
white