Then the display will show the last calibration weight used. If this is correct you
can continue by pressing the
keys to change the calibration weight value. When it is correct press the
Then display will show
ble press the
If the calibration is acceptable the display will return to normal. If an error messa-
ge is shown try calibration again as a disturbance may have prevented a success-
ful calibration.
If the problem persists then contact your dealer.
After calibration, itshould be checked to verify the calibration and linearity is
correct. If necessary repeat calibration, especially be certain the scale is stable
before accepting any weight.

11. Error code

Error
Description Resolution
codes
- - - - -
Over range
Zero
Setting
Error
A/D out of
range
46
Indicator
key. If it is not correct use the
. Place the calibration weight on the scale. After sta-
key.
Remove weight from the indicator.
If the problem persist contact your dealer for assistance.
The indicator was outside the normal zero setting range
either when it was turned on or when the
pressed.
Remove weight from the scale and try again.
Use the
key to set the display to zero value.
If the problem persist contact your dealer for assitance.
The values from the A/D converter are outside the nor-
mal range. Remove weight from the scale if overloaded,
make sure the pan is attached.
Indicates the load cell or the electronics may be faulty.
If the problem persist contact your dealer.
Model BR30
key.
key was