Removing Ameda HygieniKit Milk Collection System
from Breast(s)
• After pump is OFF, if necessary, insert a finger between
breast(s) and breast flanges(s) to break suction.
• Remove breast flanges(s) from breast(s).
Cleaning the Ameda Platinum Breast Pump

For At-Home/End-Users

After each pumping, wipe the Ameda Platinum Breast Pump
with clean, damp cloth.

For Hospitals/Rental Stations

Clean between users. Wearing disposable gloves, wipe
the Ameda Platinum Breast Pump with damp disposable
cloth using hospital approved antibacterial cleaner, such as
CaviCide
.
TM

Troubleshooting

Pump does not turn ON
• Ensure proper power cord connection to pump.
• Ensure grounded plug connection to electrical outlet.
• Ensure outlet has power.
Unable to adjust speed
• Ensure pump function is ON.
• Ensure "Check Kit" is not shown in suction display
window.
Unable to adjust suction
• Ensure pump function is ON.
• Ensure "Check Kit" is not shown in suction display
window.
Low or no suction
• Ensure "Check Kit" is not shown in suction display
window.
• Check breast flange for adequate seal against breast.
• Check that the Ameda HygieniKit Milk Collection System
is properly assembled.
• Check that adapter cap is firmly snapped onto top of
breast flange.
• Ensure silicone diaphragm is in place inside top of breast
flange.
• Ensure tubing adapter is properly fitted into tubing
adapter port on pump.
• Inspect valve for stretching or tears. If seen, replace
valve.
• Ensure valve is pushed firmly into place in lower portion
of breast flange.
• Ensure one side of tubing adapter is closed when single
pumping.

Using Timer

• Timer starts when pump
is on. Timer displays
pumping duration in
minutes : seconds.
• Timer can be reset by
pressing Timer Reset
button.
• Timer and pump
automatically turn OFF
after 60 minutes.
Turning Pump OFF
• When finished pumping,
press and hold Power
button to turn pump and
timer functions OFF.
"Check Kit" error message appears
• Ensure tubing adapter is properly fitted into tubing adapter
port on pump.
• Ensure one side of tubing adapter is closed when single
pumping.
• Check that adapter cap is firmly snapped onto top of
breast flange.
• Ensure silicone diaphragm is in place inside top of breast
flange.
• Check tubing and silicone diaphragm to ensure that parts
are free of defects.
• Check tubing is firmly attached to adapter cap.
"Service Required" error message appears
• Call your local distributor. For listing of distributors
worldwide, visit www.ameda.com.
The Ameda Platinum Breast Pump should be serviced only
by Ameda, Inc. or authorized Service Centers approved by
Ameda, Inc.

Maintenance

It is recommended the Ameda Platinum Breast Pump be
inspected at least once every year to include the following:
• Visually check pump housing for cracks, breakage, or
damage and ensure Pump Feet are present.
• Visually check for power cord damage.
• Check for proper operation of controls and buttons.
• Check there are no missing screws.
• Check cover is present on Service Port Door.
• Check Tubing Adapter Port is free from obstruction.
Quick Functional Check
The following checks/tests should be performed prior to use
by a new user:
• Visual checks for power cord damage.
• Check for abnormal noise.
• Check for proper operation of controls and buttons.
Service Port
• Do not open Service Port Door on bottom of the Pump.
• Do not connect any equipment to the service port when
operating the pump.
• Service port is to be accessed only by authorized Service
Centers approved by Ameda, Inc.
Replacing Detachable Power Cord:
• Power cord is detachable from power inlet.
• If power cord condition is in doubt, only replace with
Ameda supplied 3-conductor cord equipped with
protective ground conductor.

Technical Information

Protection Class 1, Type B
Power Supply:
Operating Input Voltage 100 VAC to 240 VAC
Operating Frequency 50/60Hz
Input Power 80VA
Fuse Rating Quick Acting (F), 2.5A, 250V, 5 x 20mm
Power Cord: Use only Ameda supplied power cord
For North America, use Ameda 24502059
Minimum Voltage Rating: 125V
Minimum Current Rating: 10A
Plug Type: IEC 60320 C13 (Hospital Grade)
Maximum Length: 10 feet (3.05m)
Cord Type: 3x18AWG, rated VW-1, 105C
Jacket Type: SJT or SJTW
Certification: UL Listed (ELBZ)
Operation: Continuous
ISO 10079-1, Intermittent Suction, Medium Vacuum
Suction: Shown in percentage of full scale suction 1%-100%;
(approximately 30mmHg to 250 mmHg/4.0kPa to 33.3kPa)
Speed: 30 to 80 cycles per minute (cpm)