LERBAKKEN 10, 8410 RØNDE DENMARK.
For safe installation of 6333B and 6343B the following must be observed. The module shall only
be installed by qualified personnel who are familiar with the national and international laws,
directives and standards that apply to this area.
Year of manufacture can be taken from the first two digits in the serial number.
IECEx Certificate
Marking
Standards:
Hazardous area
Zone 0, 1, 2, 20, 21, 22
T4: -40 ≤ Ta ≤ 85ºC
T5: -40 ≤ Ta ≤ 60ºC
T6: -40 ≤ Ta ≤ 40ºC
Terminal:
41,42,43
Uo: 30 VDC
Io: 8 mA
Po: 60 mW
Lo: 35 mH
Co: 60.8 nF
Terminal:
51,52,53
Uo: 30 VDC
Io: 8 mA
Po: 60 mW
Lo: 35 mH
Co: 60.8 nF
Revision date:
2014-06-20
IECEx Installation drawing
IECEx DEK 14.0049X
Ex ia IIC T6..T4 Ga
Ex ia IIIC Da
Ex ia I Ma
IEC60079-11:2011, IEC60079-0: 2011, IEC60079-26:2006
43
42
41
53
52
51
Version Revision
V1R0
WWW.PRELECTRONICS.COM
Non Hazardous Area
+
13
-
11
CH1
+
23
-
21
CH2
6333
6343
6333V104-ES
6333QI01
Terminal:
11,13 and
21,23
Ui: 30 VDC
Ii: 120 mA
Pi: 0.84 W
Li: 10 μH
Ci: 6.2 nF