Start & end measuring data
204
error
205
Feeler lift up error
Exceeded the measurement
206
limit.
208
LMU origin error
209
RMU origin error
210
Pattern is not set.
401
XYRf initialization error
402
X initialization error
403
Y initialization error
404
Rf initialization error
405
Rr initialization error
406
XY initialization error
407
XR initialization error
408
YR initialization error
501
Chuck origin sensor
502
Chuck close error
601
Main wheel does not rotate.
Groove wheel does not
602
rotate.
Start point and end point do not
match.
The feeler is caught on a lens during
measurement.
The lens is too thick.
Response abnormality of LMU sen-
sor
Response abnormality of RMU sen-
sor
No pattern is detected at the start of
pattern measurement.
X, Y, and Rf axes do not initialize.
X axis does not initialize.
Y axis does not initialize.
Rf axis does not initialize.
Rr axis does not initialize.
X and Y axes do not initialize.
X and R axes do not initialize.
Y and R axes do not initialize.
Response abnormality of origin sen-
sor
Chucking is not accomplished.
The wheel for the outside shape
does not rotate.
The grooving wheel does not rotate.
Contact NIDEK or your autho-
rized distributor.
Check the cleanliness of the lens
surface.
The lens is too thick to process.
Turn off power and clean the
feeler unit.
Check that a pattern is set. If
tracing is started without chuck-
ing any pattern, turn the power
switch off and on without any-
thing to be chucked.
Contact NIDEK or your autho-
rized distributor.
13