Calibration; Tds; Cell; Temperature Coefficient - YSI EC300 Manual De Operaciones

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

CALIBRATION

Calibration setup contains five sections: TDS, Cell, Temperature Coefficient, Temperature
reference, and Conductivity Calibration. To access these sections:
1. Connect the conductivity probe and cable assembly to the unit and turn the unit on. The
screen will display CELL and the cell constant of the conductivity probe.
2. Allow temperature readings to stabilize, then press CAL to enter the calibration mode; CAL
appears on the LCD. Press MODE to sequentially display the following sections:
Note: Press Enter (
advance to the next section. If there are no changes, the unit accepts the current value and
proceeds to the next section.

TDS

TDS is determined by multiplying conductivity (mS) by a TDS factor. The default factor value is
0.65. To change the TDS factor, use the ∆ ∆ ∆ ∆ and ∇ ∇ ∇ ∇ keys to adjust the value between 0.30 and
1.00. Press Enter (
display the CELL screen.

CELL

The second screen will display CELL and the current cell value. The default cell value is 5.00
and is displayed in the lower right of the screen. The unit allows a variance of ±0.50 before
displaying an error message. The cell value cannot be adjusted at this screen; calibrating
conductivity is the only way to adjust the cell constant. Press Enter (
constant to 5.00 and display the Temperature Coefficient screen.
Note: Be certain to press Enter (
the unit retains the previous cell constant and calibrates from a value that is already offset.

Temperature Coefficient

The unit uses the temperature coefficient to calculate temperature compensated conductivity.
The default value is 1.91%. To change the temperature coefficient, use the ∆ ∆ ∆ ∆ and ∇ ∇ ∇ ∇ keys to
adjust the value between 0 and 4.00%. Press Enter (
MODE to cancel the change and display the Temperature Reference screen.

Temperature Reference

The unit uses the temperature reference value to calculate temperature compensated
conductivity. The default value is 25°C. To change the temperature coefficient, use the ∆ ∆ ∆ ∆ and
∇ ∇ ∇ ∇ keys to adjust the value between 15 and 25°C. Press Enter (
press MODE to cancel the change and display the Conductivity Calibration screen.

Conductivity Calibration

1. Immerse the probe in a standard of known conductivity, preferably a standard in the middle
range of the solutions to be measured. Completely submerge the probe without touching
the sides of the calibration container. Shake the probe lightly to remove any air bubbles
trapped in the conductivity cell.
2. Allow temperature to stabilize. The message 'rAng' (range) may display briefly to indicate
unit auto-ranging; this is normal. After temperature stabilization, use the ∆ ∆ ∆ ∆ and ∇ ∇ ∇ ∇ keys to
adjust the conductivity value to that of the conductivity standard at 25°C. Press Enter (
to calibrate. The unit beeps twice to indicate a successful calibration, then automatically
switches to normal operation mode.
) to accept any values changes in each section and automatically
) to save the new value, or press MODE to cancel the change and
) to reset the cell constant to 5.00. If MODE is pressed,
) to save the new value, or press
) to save the new value, or
5
) to reset the cell
)
Tabla de contenido
loading

Tabla de contenido