5
CONNECTING THE UNIT
CAUTION: A 5-micron filter (TSD800-6) ,not included, must be installed with the unit to
ensure proper air filtration.
1
2
Items
Description
1
To Air Source
2
Air Filter (not included)
3
Power Adapter
4
Foot Switch
6
SETUP INSTRUCTIONS
Refer to Figure 1.0
6.1 Manual/Purge Dispense Cycle Setting:
1. Turn on the unit by pressing the Power button (1).
2. Press the Mode button (2) until "PURGE" appears on the Display.
3. Turn up the air pressure by rotating the Air pressure regulator knob (5) until
the desired pressure is indicated on the Display.
4. After filling the barrel or using pre-packaged adhesive, attach syringe to
receiver head assembly. Make certain that the syringe locks into place.
5. Connect the plug end of the receiver head assembly to one of the air
dispense outlet (6).
6. If vacuum "suck back" is needed, rotate the Vacuum regulator (7) counter
clockwise until the desired vacuum pressure is indicated on the Display.
7. Press and hold the Foot switch to activate dispensed cycle. (The
Manual/Purge Mode is now activated)
6.2 Automatic Dispense Cycle Setting
6.2.1 Program selection
1. Press the reset button (3) until "P:" is flashing
2. Rotate the Setup Control Knob (8) to select desired program
6.2.2 Timed Mode setting
1. Press the Mode button (2) to select "TIMED" mode
2. Press and release the reset button (3) until the colon (:) in the
"TIME:" is flashing
3
4
B
Figure 2.0
5
3. Rotate the Setup control knob (8) to set the desired dispense time
4. Depress Foot switch to activate timed dispensed cycle
6.3 Teach Mode Setting
In the teach mode, the unit will accumulate time as long as the foot switch is
depressed. This is helpful in determining dispense time required when
dispense output is unknown.
1.
Press the Mode button (2) to select "TEACH" mode
2.
Press and release the reset button (3) until the dispense time = 0.000S
3.
Press and hold down Foot switch. Dispense time will accumulate
during this time
4.
Release Foot switch when proper amount of fluid is dispensed
5.
Press the Mode button (2) to switch the "TIMED" mode
6.
The dispenser is now set to repeat this time cycle.
6.4 Multiple Shots Setting:
Up to ten separate individual or sequenced shots can be stored in the
TS350/TS355.
1.
Follow steps in section "6.2 Automatic Dispense Cycle Setting" to set
dispense time in desired programs
2.
Press both the Reset button (3) and the Power button (1)
simultaneously to activate sequence program mode
3.
Rotate the Setup Control knob to select total number of program to be
dispensed in sequence
4.
Press both the Reset button (3) and Power button (1) to exit Multiple
Shots mode
6.5 Cycle Counter
The cycle counter records the numbers of automatic dispense cycle being
activated. Up to 65,535 cycles can be recorded. This number is shown at
the lower right hand corner of the LCD. To reset the counter, press and hold
the Reset button (3) and the Mode button (2) simultaneously until the
counter shows "0".
6.6 Change Pressure and Vacuum Unit Display
1.
Press and release the reset button (3) until the colon (:) in the
"PRESSURE:" or "VACUUM:" is flashing
2.
Rotate the Setup control knob (8) to select desired unit display
6