Maintenance
The Mini Vortexer is a maintenance free product. There are no user
serviceable parts inside, only the rubber vortexing cap is replaceable.
Clean unit as often as is deemed necessary. Clean unit only when it is not
plugged into an electrical outlet. The unit can be completely wiped using
a damp cloth and a mild, non-corrosive detergent. Do not use any solvent
on the unit that may attack plastic or rubber.
Ensure all parts are thoroughly dry prior to operation.
Do not immerse the unit or the power adapter.
Trouble Shooting
Trouble
Unit does not
operate
Fluid in tube
is not forming
a visible
vortex
Flexible
vortexing
cover comes
off

Return for Repair

Important:
Transporting hazardous materials without a permit is a violation of federal
law.
Heathrow Scientific LLC will not accept any return that has not been
appropriately cleaned and decontaminated.
In the unlikely event of a return, or when damage to the unit necessitates
return, contact Heathrow Scientific LLC and obtain return authorization
before sending your piece of equipment for service. Product must be
decontaminated before transport.
Possible Cause
— Not properly plugged in
— Pressure switch not being
activated with enough
force
— Too much fluid in tube
— Cover not properly
installed
— Cover is damaged
Solution
— Plug in unit
— Verify correct
position of tube, and
press down hard
enough to activate
— Leave a minimum
of 25% of open
space in the vortex
container
— Snap cover back on
— Check cover for
damage. Replace
with spare cover
7