3B SCIENTIFIC
Geiger-Müller Counter Tube U8533430

Instruction sheet

07/07 SP/ALF
5
1. Safety instructions
If too high a voltage is applied, the life of the counter
tube will be reduced and its characteristic curve will
be altered (the gradient of the plateau will increase).
Do not apply a voltage higher than 600 V.
The mica window is only 0.03 mm thick and is sub-
jected to the external atmospheric pressure.
Caution! Do not touch the mica window.
Replace the protective cap after each use.
The glass seal of the counter tube is not resistant to
impacts and pressure.
Do not subject the tube to knocks.
Handle the counter tube especially carefully
when it is taken out of its supporting clip.
®
PHYSICS
4
3
1 Handling rod
2 BNC socket
3 Supporting clip
4 Housing
2
5 Protective cap
1
2. Description
The Geiger-Müller counter tube is a halogen self-
quenching trigger counter tube for the detection of
α-, β-, γ- and X-radiation.
The counter tube consists of a chromium steel cylin-
der that is filled with an argon-neon mixture together
with a halogen gas as a quenching agent. It is closed
at one end by a very thin mica window and at the
other end by a glass seal, both of which are gas-tight.
A thin wire (the counter wire) is arranged axially in
the cylinder, which it enters through the insulating
glass seal. The counter tube is contained within a
metal housing that is fitted with a shielded connect-
ing lead and a BNC socket. A handling rod is attached
to the housing by a detachable clamp. The mica
window is protected by a removable plastic cap.