SPECIFICATIONS
Model
Primary Outlet Flow Range
Auxiliary Outlet Flow Range
Bleed Flow
Maximum Combined
(All Outlets)
Flow
Bypass Flow (Loss of Air or
Oxygen supply)
Bypass Alarm Activation
Alarm Reset:
Alarm Sound Level:
Oxygen Concentration
Adjustment Range:
Gas Supply Pressure:
Mixed Gas Stability:
Connection Types:
NOTE: All flow-rate values are as measured from an Oxygen flowmeter (uncorrected).
DIMENSIONS:
Operating Temperature Range:
866.4.MAXTEC
MicroMax High Flow
R203P14
15 - 120 l/min
With both supply pressures at 50 psi (3.4 bar) with BLEED closed
2 - 100 l/min
With both supply pressures at 50 psi (3.4 bar) with BLEED open
13 l/min or less
at 50 psi (3.4 bar)
≥ 120 l/min
> 85 l/min
50 psi (3.45 bar)
13-25 psi
0.9-1.7 bar
When pressure differential is 6 psi (0.4 bar) or less.
≥ to 80 db at 1 ft (0.3 m)
21 - 100%
30 - 75 psi (2.1 - 5.2 bar)
Air and Oxygen within 10 psi (0.69 bar) of each other
±1% Oxygen
DISS Type - Air & Oxygen Inlets & Outlets and / or NIST Type - Air &
Oxygen Inlets
(without fittings)
Height:
Weight:
Shipping Weight:
60 psi
(4.14 bar)
16-24 psi
1.1-1.65 bar
4.9 in
Depth:
Width:
2.3 in
4.1 in
2.29 lbs
2.95 lbs
59°F to 104°F (15°C to 40°C)
v
MicroMax Low Flow
R203P13
3 - 30 l/min
0 - 30 l/min
3 l/min or less
at 50 psi (3.4 bar)
≥ 30 l/min
> 45 l/min
50 psi
(3.45 bar)
18-22 psi
1.2-1.5 bar
(12.5 cm)
(5.7 cm)
(10.4 cm)
(1.04 kg)
(1.34 kg)
WWW.MAXTEC.COM
60 psi
(4.14 bar)
16-24 psi
1.1-1.65 bar