INSTALLATION
The installation process consists of the following stages:
Preparation
Please read the following sections thoroughly before proceeding with the installation.
PREPARATION
Ensure that the sensor is positioned so that it is focused on the target only.
DISTANCE AND SPOT SIZE
The size of the area (spot size) to be measured determines the distance between the sensor and
the target. The spot size must not be larger than the target. The sensor should be mounted so
that the measured spot size is smaller than the target.
AMBIENT TEMPERATURE
The sensing head withstands up to 60°C, 120°C, or 180°C, without cooling, depending on the
choice of model. See "Model Numbers" for more information.
Avoid thermal shock. Allow 20 minutes for the unit to adjust to large changes in ambient
temperature.
ATMOSPHERIC QUALITY
Smoke, fumes, dust or steam can contaminate the lens and cause errors in temperature
measurement. In these types of environment the air purge collar should be used to help keep the
lens clean.
INTERFERENCE FROM MOVEMENT
The low-noise sensing head cable on -HA models is resistant to interference caused by movement.
The sensing head may be mounted on moving machinery such as robot arms without affecting the
accuracy of the measured temperature.
ELECTRICAL INTERFERENCE
The PyroMini is tested to industrial standards for electromagnetic compatibility (EMC) as shown in
Specifications at the beginning of this manual.
To minimise electromagnetic interference or 'noise', the sensor should be mounted away from
motors, generators and such like.
1 4
Mechanical installation
Sensor
Target greater
than spot size
Target smaller
than spot size
Background
Electrical installation
CORRECT
INCORRECT