DeFelsko PosiTest AT-M Manual De Instrucciones página 7

Ocultar thumbs Ver también para PosiTest AT-M:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Step 4b: Pull-Off Test
Step 4b: Pull-Off Test
USB port
AC Adapter Port
(Charges batteries
and powers instrument)
Quick Guide
(1) Connect the actuator to the dolly
(2) Power-up with the
(3) Verify measurement units, dolly size and pull rate
(4) Ready the instrument with the
(5) Test with the
1.
Place the actuator assembly over the dolly head and attach the quick coupling
to the dolly by reaching through the holes in the actuator assembly and lifting the
quick coupling. Release the quick coupling when the dolly head is completely
engaged.
2.
Press the
ment will power-down after 5 minutes of no activity or by holding the
ton for 2 seconds.
3.
Check settings:
(a) Verify displayed measurement units. Change with the
necessary.
(b) Verify dolly size with the
(c) Verify pull rate with the
following user selectable rates are available:
Dolly Size
10 mm
14 mm
20 mm
50 mm
(PosiTest AT-A Automatic)
LCD
button
button to power-up the instrument if necessary. The instru-
PSI Rates
125, 200, 400, 600, 725
60, 100, 200, 300, 360
30, 50, 100, 150, 180
5, 8, 16, 24, 30
Page 8
Quick
Coupling
button
button
button and change if necessary.
button and change if necessary. The
1.00, 2.00, 3.00, 4.00, 5.00
0.40, 0.70, 1.40, 2.00, 2.50
0.20, 0.30, 0.70, 1.00, 1.20
0.04, 0.08, 0.12, 0.16, 0.20
Hose
Actuator
Handle
Actuator
Assembly
button if
MPa Rates
but-
Tabla de contenido
loading

Este manual también es adecuado para:

Positest at-a

Tabla de contenido