bewell connect MyTens Manual De Instrucciones página 29

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Power
3 A/m
frequency
(50Hz/60Hz)
magnetic field
IEC
61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
GUIDANCE AND MANUFACTURE'S DECLARATION – ELECTROMAGNETIC IMMUNITY
The BW-TSX / TS1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BW-TSX / TS1 should assure that it is used in such an
environment.
IMMUNITY
IEC 60601 TEST
LEVEL
TEST
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
Radiated RF IEC
MHz
61000-4-3
3 V/m
80 MHz to 2.6
GHz
56
BewellConnect - BW-TSX/BW-TS1 - User Manual - 12-2018_v2
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
COMPLIANCE
ELECTROMAGNETIC
LEVEL
ENVIRONMENT - GUIDANCE
3 V RMS outside
Portable and mobile RF com-
the ISM band,
munications equipment should
6 V RMS in the
be used no closer to any part of
ISM and amateur
the BW-TSX / TS1 including cables,
radio bands 10
than the recommended separation
V/m
distance calculated from the equa-
tion applicable to the frequency of
the transmitter.
Recommended separation
distance
d =
1,2 P √
d =
1,2 P 80 MHz to 800 Mhz
d =
2,3 P 800 MHz to 2,5 Ghz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the BW-TSX / TS1 is used
exceeds the applicable RF compliance level above, the BW-TSX / TS1 should be observed
to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating the BW-TSX / TS1.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATIONS EQUIPMENT AND THE BW-TSX / TS1
The BW-TSX / TS1 is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the BW-TSX / TS1 can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the BW-TSX / TS1
as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum
output power
Separation distance according to frequency of transmitter (m)
of transmitter
(W)
150 KHz to 80 MHz
0,01
d =
1,2 P √
BewellConnect - BW-TSX/BW-TS1 - User Manual - 12-2018_v2
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,a should be less than
the compliance level in each
frequency range.b Interference
may occur in the vicinity of
equipment marked with the
following symbol:
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d =
d =
1,2 P √
2,3 P √
57
Tabla de contenido
loading

Productos relacionados para bewell connect MyTens

Este manual también es adecuado para:

Mytens proBw-tsxBw-ts1

Tabla de contenido