25 Recommended calibration intervals 
SDT recommends annual recalibration
This periodicity is coherent with the observed long‐term stability of SDT electronic equipment's. 
However 
For new equipment leaving the factory the initial period is extended to 15 months 
instead of 12  (to avoid any possible discrimination between users due to transportation time, customs clearance, 
etc.)   
SDT recognizes that each customer may have their own Quality Assurance Standards and internal requirements.  If 
a customer has established a Quality Assurance Program that includes a documented procedure for measurement 
deviations,  
and if an instrument and the relevant sensor show acceptable deviations  
and do not show signs of physical damage, then calibration periodicity could be extended to two years. 
Third party organizations having more authority (Class, or State Administrations) may 
have defined their own regulations for example 6 months, or 2 years calibration interval:  
by evidence, local legally applicable regulations supersede SDT rules. 
For these reasons, SDT Calibration Certificates do no more mention a "calibration due date",  
but simply the "last calibration date". 
For user convenience, the date of the last calibration of an SDT200 or 270 instruments 
can also be found in the menu screens. 
This text refers to SDT document DC.QUAL.005
 
                                                                  
1
 This document is part of SDT's quality management system (ISO‐9001 certified). 
 
 
of SDT instruments and annual verification of sensors. 