CARE AND MAINTENANCE
Proper care and maintenance of your blood pressure monitor will help ensure long
life for the product; improper use of the monitor may void the included warranty.
Monitor Care and Maintenance
•
Do not use any liquids on the monitor; use a soft, dry cloth to clean the monitor
as needed
•
Storage outside of stated storage temperature may result in measurement
error or device malfunction; storage environment temperature is: -4°F – 122°F
(-20°C – 50°C); humidity: 15% to ≤85%Rh (non-condensing)
•
Remove the batteries if the unit will not be used for an extended period of time
•
Do not attempt to disassemble your monitor; disassembling your monitor may
void the manufacturers warranty
•
Do not subject the monitor to strong shocks; take care not to drop the monitor
Cuff Care and Maintenance
•
The cuff should not be submerged or washed in a washing machine, but may
be spot cleaned with a mild-detergent as needed
•
never attempt to iron or dry the cuff; if the cuff becomes wet, allow the cuff to
air dry completely before use or storage
•
To disinfect the cuff, wipe the inside (skin contact side) of the cuff using a cloth
lightly moistened with 70% - 90% ethyl rubbing alcohol and allow to air dry;
it is suggested to disinfect the cuff frequently when used in clinical sittings or if
the cuff is contaminated in any way.
18 • English