Tabla De Contenido; Specifications - Hach 2100AN IS Manual Básico Del Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Table of contents

Specifications

Specifications are subject to change without notice.
Specification
Details
Measurement method
Nephelometric
Regulatory
Meets ISO 7027, DIN EN 27027, DIN 38404 and NFT
9033
ASTM D7315 - Standard Test Method for
Determination of Turbidity Above 1 Turbidity Unit (TU)
in Static Mode
ASTM D6655 - Standard Test Method for
Determination of Turbidity Below 5 NTU in Static Mode
Light source
Light-emitting diode (LED) at 860 ± 30 nm
FNU, FAU, NTU, EBC, Abs (absorbance), %T (%
transmittance) and two user-defined units
Standard operation
on page 11
Maintenance
on page 16
Troubleshooting
on page 17
Specification
Details
Range
FNU (manual range): 0-0.999, 0-9.99, 0-99.9,
0-1000
FNU (auto range): 0-1000
FAU (manual range): 20-99.9, 20-10,000
FAU (auto range): 20-10,000
0-99.9, 0-10,000
NTU (Ratio on, auto range): 0-10,000 auto decimal
NTU (Ratio off): 0-40
EBC (Ratio on, manual range): 0-0.999, 0-9.99,
0-99.9, 0-2450
EBC (Ratio on, auto range): 0-2450 auto decimal
EBC (Ratio off): 0-9.8
Absorbance (manual range): 0-0.999, 0-2.00
Absorbance (auto range): 0-2.00
Transmittance (%): 1.0-100
1, 2, 3
4
Accuracy
FNU
: ±2% of reading plus 0.01 FNU from
0-1000 FNU
FAU: ±10% of reading from 20-10,000 NTU
4
: ±2% of reading plus 0.01 NTU from
0-1000 NTU, ±5% of reading from 1000-4000 NTU,
±10% of reading from 4000-10,000 NTU
Absorbance: ±0.005 Abs from 0-1 Abs at 860 nm
Transmittance: 0.12% T from 10-100% T at 860 nm
Resolution
Turbidity: 0.001 FNU/NTU/EBC
Absorbance: 0.001 Abs
Transmittance: 0.1% T
Repeatability
±1% of reading or 0.01 FNU, whichever is greater
(under reference conditions)
Response time
Signal averaging off: 6.8 seconds
Signal averaging on: 14 seconds (when
10 measurements are used to calculate the average)
English 3
Tabla de contenido
loading

Este manual también es adecuado para:

2100an

Tabla de contenido