Operation Instructions; Maintenance - Health O Meter 1101KL Manual De Operación

Balanza de gran capacidad
Idiomas disponibles

Idiomas disponibles

OPERATION INSTRUCTIONS

1.
Press the ON/OFF button to turn the scale on.
2.
Wait until "0.0" appears on the digital display.
3.
Ask the patient to step on the scale. The display will blink "
person is determined.
4.
If you wish to reweigh without asking the patient to step off and to step on the scale
again, press the REWEIGH button.
5.
Ask the patient to step off the scale. The digital display should continue to read the
person's weight until REWEIGH or ZERO button is pressed.
NOTE: After 2 minutes of inactivity, the scale will turn itself off to conserve power. To restart
the scale press the ON/OFF button.

MAINTENANCE

GENERAL
This section provides instructions for maintenance, cleaning, troubleshooting and operator
replaceable parts for the High Capacity Stand-On Scale Model 1101KL. Maintenance
operations other than those described in this section should be performed by qualified service
personnel.
MAINTENANCE
Before first use and after periods of non-use, check the scale for proper operation and
function. If the scale does not operate correctly, refer to qualified service personnel.
1.
Check overall appearance of the total scale for any obvious damage, wear and tear.
2.
Inspect AC adapter for cord cracking or fraying or for broken or bent prongs.
CLEANING
Proper care and cleaning is essential to ensure a long life of accurate and effective operation.
Disconnect the scale from the AC power source.
1.
Clean all external surfaces with a clean damp cloth or tissue. Mild soap and water
solution may be used. Dry with a clean soft cloth.
2.
Do not immerse the scale platter or pillar assembly into cleaning or other liquid solution.
3.
Do not use Isopropyl Alcohol or other solutions to clean the display surface.
------
------
6
" until the weight of the
loading