After fine alignment of both jaws:
▷ Start the matching process by clicking on the
The defined points for the fixator registration are stored in the software.
Fig. 18
The registration of the fixator is completed. From now on, all subsequent jaw scans of patient cases are
automatically assigned.
The more accurate the points are set, the more precise the assignment. Small deviations are per-
mitted, and will be corrected in the matching process!
When the assignment was created incorrectly:
▷ Select
Scanner Options > Reset jaw alignment
▷ Click the
Start artikulation
▷ Click on
Point alignment
▷ Start the registration process by clicking on
Continue
to reset the incorrect registration.
button.
.
OK
I N S T A L L A T I O N
button.
.
E N
45