higher measurement range to a
lower measurement range.
• Before using the device, check the
housing and the measurement
cables for any damage.
• Do not conduct measurements in
a damp environment. Workplace,
hands, shoes and floor must be dry.
• Before opening the housing all
measurement leads are discon-
nected from the device.
2. Description, technical data
Passive analog multimeter with a
rotary switch for selecting the mea-
surement ranges and a scale with
mirrored background for parallax-
free readings. The device is extreme-
ly robust in terms of load capacity
and is equipped with excellent over-
load protection due to its two anti-
parallel diodes as well as moving coil
movement, which is not sensitive to
external electromagnetic fields. The
safety connection sockets offer pro-
tection against accidental touch con-
tact. The robust plastic housing and
the spring-connected bearing jewels
of the movement guarantee protec-
tion against damage caused by me-
chanical stress.
Measurement ranges:
Voltage measurement:
DC
Meas. range
Int. resistance
100 mV
2 kΩ
1 V
20 kΩ
10 V
200 kΩ
30 V
600 kΩ
100 V
2 MΩ
300 V
6 MΩ
600 V
12 MΩ
AC
Meas. range
Int. resistance
10 V
66,7 kΩ
30 V
200 kΩ
100 V
667 kΩ
300 V
2 MΩ
600 V
4 MΩ
Current measurement:
DC
Meas. range
Voltage drop
50 µA
100 mV
1 mA
500 mV
10 mA
500 mV
100 mA
500 mV
1 A
190 mV
AC
Meas. range
Voltage drop
3 mA
1.5 V
30 mA
1.6 V
300 mA
1.6 V
3 A
1.8 V
Resistance measurement:
Rotary
Meas. range
switch
and scale center
Ω x 1
1 Ω...35 kΩ...5 kΩ
Ω x 10 10 Ω...350 kΩ...50 kΩ 4.5 mA
Ω x 100100 Ω...3.5 kΩ...500 kΩ 0.45 mA
Accuracy: Class 2,5
Influencing variables and nominal
operating ranges:
Temperature 0 – 40° C:
± 1% / 10 K for DC; ± 2.5% / 10 K
for 100 mV/50 µA DC; ± 1.5% / 10 K
for AC
Frequency (30 Hz...1 kHz): ± 2.5%
Reference conditions:
7
Max.
meas.
current
45 mA