Error Codes; Corrective Action - smartheart 01-508 Manual De Instrucciones

Ocultar thumbs Ver también para 01-508:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

ERROR CODES

Code
E 1
The cuff is not secure/nottight
enough.
E 2
The cuff is wrapped too
tightly.
E 3
The cuff pressure is overinflated.
Excessive movement was detected
E 10
during measurement.
E 11
An error occurred during
E 20
measurement and the results are
E 21
inaccurate.
Eexx
A calibration error has occurred.
OUT
Out of measurement range.
IF AN ERROR CODE PERSISTS OR IS NOT REMEDIED,
PLEASE CONTACT CUSTOMER CARE AT 866-326-1313.
26 • ENGLISH
Meaning

Corrective Action

Re-apply the cuff and attempt new
measurement.
Re-apply the cuff and attempt new
measurement.
Relax for several minutes and attempt
new measurement.
Remain still and attempt new
measurement.
Relax for several minutes and attempt
new measurement.
A calibration error has occurred.
Result is out of range; relax for a
moment and attempt new measument.

Capítulos

Tabla de contenido

Solución de problemas

loading

Tabla de contenido