5. To adjust the voltage or current scaling for other than zero to full scale, press and hold the MODE key while
pressing the FLOW key. This display will show "LO" for 2 seconds and then a flow rate. Use the UP/DOWN (▲, ▼)
arrow keys to set the minimum control level. Press the FLOW key again. The display will show "HI" for 2 seconds
and then a flow rate. Use the UP/DOWN (▲, ▼) arrow keys to set the maximum control level. Press any other key
to exit. The same scaling will be used for both input and output levels.
NOTE:
Tubing size should be selected before adjusting the voltage or current scaling.
Remote START/STOP
Use of the remote START/STOP is optional. It can be used to start the pump in
any mode by closing the input. In dispense or copy modes, only a momentary
remote START/STOP closure is needed to start or stop a dispense or copy. In mA
or V modes, there must also be a non-zero signal present. Once the remote
START/STOP has been closed to start the pump, opening it will stop the pump. In
mA and V modes, it is best to wire your START/STOP contacts in series with the
mA or V control signal to insure the pump will always be off when the
START/STOP contacts are open.
PRIME
CW/CCW
RED/YELLOW
START/STOP
YELLOW
INPUT
0-20mA; 4-20mA
ORANGE
INPUT
0-10V
BROWN
OUTPUT
0-20mA; 4-20mA
OUTPUT
0-10V
14
NOTE:
BLUE
6
GREEN
4
10
8
12
WHITE
3
11
5
13
BLACK
1
15
16
7
VIOLET
9
17
RED
2
18
4
Colors are those of Remote Cable,
Catalog number 77300-32.
GREY
TACH OUTPUT
TAN
PINK
MOTOR RUNNING
N.O. CONTACT
RED/GREEN
MOTOR RUNNING
N.C. CONTACT
RED/BLACK
N.C.
N.C.
N.C.