DeFelsko PosiTector 6000 Guía Rápida página 3

Ocultar thumbs Ver también para PosiTector 6000:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Button Functions- Normal Operation
Adjust Down
Access the Menu
To disconnect a probe from a body, power-down the
instrument and slide the plastic probe connector
horizontally (in the direction of the arrow) away from the
body. Reverse these steps to attach a new probe.
When powered-up the PosiTector automatically
determines what type of probe is attached and does a
self-check.
Additionally, the PosiTector accepts a number of probe types
including magnetic and eddy current coating thickness, surface
profile, environmental and ultrasonic wall thickness probes.
For the latest information on probe interchangeability,
see www.defelsko.com/probes
Calibration, Verification, and Adjustment
Calibration, Verification, and Adjustment
The PosiTector 6000 non-destructively measures the thickness of
coatings on metals.Three steps ensure best accuracy...
1. Calibration: typically done by the manufacturer or qualified lab.
All probes include a Certificate of Calibration.
2. Verification of Accuracy: as done by the user on known
reference standards such as the included plastic shims or
optional coated thickness standards.
3. Adjustment: Adjustment, or Cal Adjustment is the act of
aligning the Gage's thickness readings to match that of a known
sample in order to improve the effectiveness of the Gage on a
specific surface or in a specific portion of its measurement
range. 1-point or 2-point Cal adjustments are possible.
Probes
Probes
2
Adjust Up
Restore brightness
after dimming
(Advanced only)
Tabla de contenido
loading

Tabla de contenido