FIGURE 4 CARTRIDGE ASSEMBLY
3. Select tubing and place in Cartridge (see Figure 6B). If using a Microbore Tubing Set (see Figure 5), place the Collar
up against the Tubing Retainer (see Figure 6C) on both sides of the Cartridge.
4. Attach the Hook Leg of the Cartridge onto one of the Support Rods while holding the tubing between the Cartridge
legs and against the Tubing Retainer (see Figure 6D).
5. Swing the Cartridge with the tubing over the rotor and push down on the Tab (see Figure 4) until the Latch snaps
closed. (See Figure 6E.) Note: Depress the Cartridge as required in the location indicated in Figure 6E to facilitate
latching and unlatching.
6. Lightly pull the tubing at the outlet and inlet to remove the slack when using continuous tubing. No slack removal is
required when using a Microbore Tubing Set.
7. Adjust the occlusion setting (see OPERATION section). For a nominal setting with MASTERFLEX tubing, turn ad-
justment screw until inside edge of the wedge is aligned midway between #3 and #4 on the label.
6A
6D
FIGURE 6 TUBING AND CARTRIDGE LOADING
FIGURE 5 MICROBORE TUBING SET
6B
6
6C
6E