5 Use the M4 screw provided with the bridge kit to secure the bridge piece to the thermoregulator at
the heater end of the thermoregulator, do not over tighten it. Thumb screw 1 secures both the unit
and the bridge to the bath. Thumb screw 2 secures both the cover and the bridge to the bath. Thumb
screw 3 is fitted at Techne and secures the cover to the bridge. A bung is provided to seal off the
thermometer hole in the bridge piece if a thermometer is not used. It is necessary to fit this bung to
reduce heat losses and prevent steam or other vapours getting to the thermoregulator.
6 If you do not want to use a bridge piece to secure the unit to a bath then you must use a clamp. The
clamp bracket fits between the pump legs in the slot provided. Slide the bracket into the slot and use
the two screws in the clamp kit to secure the clamp to the unit.
Slot for clamp
7 If the pump has been set correctly for external circulation, suitable hoses should be fitted to the outlet
and return pipes. A suitable hose must be capable of withstanding both the temperature of operation
and the liquid being used. Always securely clip the hoses in place.
Hose material
Allowable temperature range
PVC
10°C to 60°C (50°F to 140°F)
Silicone
-40°C to 200°C (-40°F to 362°F)
Viton
-20°C to 250°C (-4°F to 482°F)