9. With test strip still in meter, touch
Sample Tip of test strip to top of
drop of control solution. Allow
drop to be drawn into test strip.
Remove test strip from drop when
meter beeps.
10. Dashes appear across the Display to
show meter is testing.
Note: If meter does not beep and begin testing soon
11. Compare meter result to
Control Test range printed on
test strip vial label for level of
control solution you are using.
If result is in range, System can
be used for testing blood. If
result does not fall within range,
repeat test using a new test strip.
Note: Control Test result shows the
TMX_AIR_OB_GutsENG_REA4TVH03r53.indd 20
TMX_AIR_OB_GutsENG_REA4TVH03r53.indd 20
after drawing up sample, release and discard
test strip. Repeat test with new test strip. If
problem persists, see Troubleshooting.
Control Symbol in the Display.
Control Solution
Does not represent actual
Control Test ranges.)
20